Abstract
Potential errors in XRF analysis are discussed and investigated with specific reference to the analysis of ilmenite samples. A selection of ilmenite samples was chosen and prepared using two XRF preparation techniques: pressed powder pellet and unroasted fusion bead. The merits of the two methods have been discussed in this paper for ilmenite analysis by XRF. Recommendations have been made regarding the use of pressed powder pellets especially with regard to the mineralogical effects that were evidenced during this investigation. The fusion bead technique was found to be far more accurate and able to meet short turnaround requirements provided the calibration methodology used is specific to the grade of samples analyzed.