RI 6009 Calibration Studies Of Three Portable Seismographs

The National Institute for Occupational Safety and Health (NIOSH)
David E. Fogelson
Organization:
The National Institute for Occupational Safety and Health (NIOSH)
Pages:
25
File Size:
4104 KB
Publication Date:
Jan 1, 1962

Abstract

One model each of three different makes of portable displacement seismographs was calibrated on a shaking table capable of 5,000 pounds of force, Calibration tests were conducted in the frequency range of 5 to 200 c.p,s, and in the amplitude range of 0,002 to 0.05 inch peak-to-peak, The seismographs were tested under two mounting conditions: (1) The feet of the instrument were bolted to the shaking table, and (2) the feet were resting on emery cloth affixed to the table, The results of these tests show that, in the frequency range of 5 to 40 c.p,s, and for vibration levels less than 0.2 g, the magnification for all three instruments is constant with input level but varies slightly with frequency. For vibration levels greater than 0.4 g. and for frequencies greater than 40 c.p.s., the magnification of all three instruments shows large variations with frequency and amplitude of table motion.