RI 6212 A Metal Diaphragm Apparatus For Measuring Vapor Pressures - Vapor Pressure Of Arsenic (III) Oxide

- Organization:
- The National Institute for Occupational Safety and Health (NIOSH)
- Pages:
- 23
- File Size:
- 8943 KB
- Publication Date:
- Jan 1, 1963
Abstract
A metal diaphragm pressure relay system was constructed to measure vapor pressures of the metal halides, oxyhalides, and other substances at temperatures up to 700° or 800° C. The vapor pressure of arsenic (III) oxide (As2O3) was determined to verify the reliability of the apparatus for such measurements. The data for the liquid AS2 Os (2820 to 4830 C) are represented by the equation, log Pmm = 6.875 - 2,929/T, from which the normal boiling point was calculated to be 461° C. The heat of vaporization calculated from this equation was 13.40 ±0.09 kcal/g mole As4OS and is essentially constant over the interval studied. The melting point of the octahedral arsenic (III) oxide was determined to be 278° C from the inter-section of the solid and liquid vapor pressure curves, A thermal analysis apparatus is also described by which melting and transition temperatures could be checked by a method independent of vapor pressure data. The melting points of the monoclinic and octahedral forms were determined to be 314.4 ±1.4° C and 278 ±5° C, respectively.
Citation
APA:
(1963) RI 6212 A Metal Diaphragm Apparatus For Measuring Vapor Pressures - Vapor Pressure Of Arsenic (III) OxideMLA: RI 6212 A Metal Diaphragm Apparatus For Measuring Vapor Pressures - Vapor Pressure Of Arsenic (III) Oxide. The National Institute for Occupational Safety and Health (NIOSH), 1963.